Do air-tightness testers with a resolution of 0.01 Pa really exist? Check these four things before placing an order.

existAir tightness test equipmentIn the marketplace, you often see advertisements like this:

highly accurateAirtightness Tester, with a resolution of 0.01 Pa!
Ultra-high precision, down to 0.01 Pa!
Import-grade precision—domestic equipment can achieve it too!

The smaller the number, the more it appears to reflect technical capability—and the more likely it is to lead to mistakes when selecting a model.

So.With a resolution of 0.01 PaAir tightness testerDoes it really exist?What exactly does 0.01 Pa represent?If the resolution is 0.01 Pa, does that mean the device's measurement accuracy is also 0.01 Pa??

I. First, the answer: It exists, but it can’t be used this way.

With a resolution of 0.01 PaAirtightness TesterExists.Renowned worldwideLeak TesterFor example: According to the official specifications for the ATEQ F620, the maximum resolution in the 0–50 Pa range is 0.01 Pa, and it can reach 0.001 Pa in a specific 20 Pa configuration.

Do air-tightness testers with a resolution of 0.01 Pa really exist? Check these four things before placing an orderIllustration
French brand ATEQ F620

On the same instrument, when the measurement range is set to 500 Pa, the resolution is 0.1 Pa; when the range is set to 5,000 Pa, the resolution is 1 Pa.TheResolution varies with the measurement range; a single minimum value does not represent the entire instrument.

II. Three Terms That Are Often Used Interchangeably

  • Number of digits displayedIt refers to how many decimal places the screen can display.
  • resolution (of a photo)It is the smallest change that the instrument can detect.
  • AccuracyIt refers to how close the measurement result is to the true value.

Let me give you an example.

If an instrument has a resolution of 0.01 Pa and a total error of ±0.5 Pa, and the screen displays 10.01 Pa, the actual value is likely between 9.51 and 10.51 Pa.

Resolution refers to the markings on a ruler; the density of the markings does not necessarily indicate the ruler’s accuracy. Even a ruler with the densest markings will still give inaccurate measurements if it is bent.

Along with accuracy, there are two other metrics:

  • repeatableCheck whether the results of repeated measurements under the same conditions are consistent,
  • stabilitySee if the instrument can maintain its performance over a long period of time.

These four indicators are interrelated, and none can replace the others.

III. Why Relying Solely on Resolution Can Lead to Errors

Leak testing is a series of measurement processes, and sensors are only one part of it. Other factors include the stability of the pressure-regulating valve and the gas source, the sealing of the gas lines and fixtures, product volume and housing deformation, gas-filling equilibration time, and zero-point drift,Fluctuations in each ring may exceed 0.01 PaThe

Formula for the ideal gas law in airtightness testing
Equation of State for an Ideal Gas: PV = nRT

The effect of temperature is the most obvious. Based on the ideal gas law, a 1°C change in temperature within a sealed chamber results in a pressure change of approximately 0.34 percent; at near-atmospheric pressure, this amounts to over 300 Pa—more than 30,000 times 0.01 Pa. Therefore, professional equipment must use differential pressure configurations and reference chambers to compensate for such interference, rather than relying solely on the sensor’s resolution to withstand it.

IV. Reference 1: How to Enter Parameters for the ATEQ F620

Measurement RangeAccuracyMaximum Resolution
50Pa±(1% of reading + 1% of full scale)0.01 Pa
500 Pa±(1% of reading + 0.2% of full scale)0.1Pa
5,000 Pa±(1% of reading + 0.2% of full scale)1Pa

Rdg is the actual reading, and FS is full scale.. The accuracy of the 50 Pa range applies to laboratory calibration conditions and cannot be assumed to be achieved by all standard configurations. The definition of accuracy includes bias, repeatability, and hysteresis; refer to the official documentation for the corresponding version for details.

This document is worth referring to becauseEach resolution is followed by the measurement range, error composition, and calibration conditions.. With this, engineers can determine exactly what performance levels the equipment can achieve under various conditions—without having to guess.

V. Reference 2: Jingcheng Engineering JC-C10005 What do you think?

Parameter ItemsPublic Parameters
Differential pressure range±5,000 Pa
resolution (of a photo)0.1 Pa, 0.001 mL/min
Differential Pressure Accuracy0.075% F.S.
Pressure accuracy0.1%F.S. / 0.2%F.S.
Time Range0–999.9 s
communication interfaceRS485, RJ45, optional RS232
communication protocolsModbus RTU, Modbus TCP

The above information is taken from the official website's product page; please refer to the technical specifications of the actual device for details.

Differential Pressure Leak Tester
China Brand Jingcheng Engineering JC-C10005 Differential Pressure Airtightness Tester

One more thing:%F.S. is the fixed error band; %Rdg varies with the reading. The accuracy of the two devices cannot be directly compared; they must be converted to the same measurement range and the same reading to be comparable.

VI. How Are Market Metrics Presented?

In China's airtightness testing market,There are examples such asShenzhen Jingchenggongke Technology Co., Ltd.(JCGK) This kind of long-term dedicationAirtightness TesterSome are manufacturers, while others are companies that have expanded into this field from automation equipment, custom-built machinery, and tooling and fixtures.The
There’s nothing wrong with working across disciplines; integrators are familiar with PLCs, robotic arms, and production lines. However,airtightness meter(used form a nominal expression)The core components are sensor measurement, gas flow control, error analysis, long-term calibration, and software algorithms; these aspects require a deep understanding of metrology.The

There are five common types of packaging:

① Focus only on resolution, not accuracy;
② Specify only the accuracy, not the measurement range;
③ Only mentions good reproducibility without specifying the test conditions;
④ It only mentions that high-precision sensors are used, without specifying the overall system error;
⑤ Only a single test result is provided; no continuous record is provided.

Each one looks like a specification on its own, but when taken together, they make it impossible for procurement professionals and engineers to make a decision.

VII. Six Questions That Let the Parameters Speak for Themselves

Asking the manufacturer questions is more effective than looking at brochures.

concernAsk why
Is the accuracy %FS or %Rdg?Is the determination error a fixed value, or does it vary with the reading?
Which range does this correspond to?The same instrument has different accuracy levels depending on the measurement range.
Does it include repeatability and hysteresis?Accuracy that includes only the bias is only half the accuracy.
Is there a standard leak verification record?Pressure changes are meaningful only when they correspond to a known leak rate.
Can the test be repeated consecutively?Examine the data's dispersion and zero-point drift
Can this be tested on my products and tooling?Laboratory data is not the same as production line data

When verifying, focus on three key points:

  1. Use a calibrated standard orifice to establish a correlation between the pressure changes on the screen and a known leak rate;
  2. Conduct continuous testing under the same conditions to determine whether the results are consistent and whether there is any zero-point drift;
  3. Finally, run it again in the actual production environment.

A fluctuation of 0.01 Pa observed on the screen does not constitute evidence; it is only considered evidence if it can be reproduced, correlated, and used to distinguish between good and defective products.

The selection process follows this same logic:First, determine the detection principle(Gauge pressure, differential pressure, flow rate, volume, etc.), then set the measurement range, verify the accuracy, check the repeatability and stability, and finally evaluate the entire system comprising the gas lines, fixtures, volume, temperature, cycle time, and automated controlsThe

VIII. Domestic Substitution: Relying on Validation Under Identical Operating Conditions

inletLeak test equipmentWhen it comes to replacing products like ATEQ, COSMO, and Uson, relying on slogans alone won’t work. The approach is to identify the model numbers and testing requirements of the original equipment and have domestic manufacturersAir tightness testing equipmentexistSame product, same tooling, same pressure, same acceptance criteriaCompare the following and verify these six items:Are the results consistent? Can good and defective products be distinguished? Does the standard leak test meet the requirements? Do repeatability and stability meet the requirements? Does the cycle time meet production needs? Are maintenance and automation integration convenient?. Through routine air tightness testing and suitable differential pressure applications, the domestic brand Jingcheng Engineering can gradually build up its capability to serve as a substitute in this way.

Conclusion

Returning to the original question:Detectors with a resolution of 0.01 Pa do exist; the specifications for the imported ATEQ F620 model provide a clear reference.The
But the more important question is:What measurement range does it cover? What is its accuracy? Does it include repeatability and hysteresis? Has it been calibrated? Can it reliably detect target leaks in actual production equipment?The

The same applies to domestically manufactured equipment. Jingcheng Engineering (JCGK)’s JC-C10005 With the differential pressure range, resolution, and accuracy now disclosed, you can select equipment, adapt tooling, and conduct actual testing based on the product’s inspection requirements. Specific performance specifications are subject to the corresponding configuration and technical specifications.

The market can compete on price, service, and delivery, and it can also compete on technology, but it shouldn’t be about who’s better at embellishing specifications.

Resolution can be very low, but the definition of error must be unambiguous; marketing specifications can be impressive, but detection capabilities must stand up to verification.

Source: ATEQ F620 Official Technical Specifications; Jingcheng Engineering JC-C10005 Official product documentation.